BD BodyGuard™

The BD BodyGuard™ infusion pump is a small, lightweight ambulatory infusion pump which offers flexibility and reliability. Effortless transition between hospital and home care requirements is the core of the BD BodyGuard™ infusion pump.
Dimensions and weight
Weight (with standard battery) 410 g
Pump only (H/W/D) 112 x 89 x 40 mm

Wide flow rate range
0.1 - 1200 ml/h
Piggyback (Secondary) 0.1 - 1200 ml/h
Bolus 0 - 1200 ml/h
KVO Rate 0 (OFF) - 20 ml/hr
Flow rate accuracy* ± 5 %

Volume to be infused
0.1 - 9999 ml

Patient safety
Ultrasonic Air Sensor Air bubble size 0.04 – 2 ml
Occlusion Pressure 100-1500 mmHg (10 mmHg increments)

Programmable infusion time (Dose over time)
100 hrs

BD bodyGuard
Complies with the current version of the following standards
IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 60601-1-11, IEC 60601-1-12, IEC 60601-2-24

Protocols library
Fully programmable drug library, with automatic calculation of drug concentration programmable via PC with BodyComm™ software**
Up to 128 drugs referred to as protocols
2000 user events log

Power supply
110-240 VAC, 50/60 Hz. 0.3A max
Battery specifications.
Rate (mL/h) Approx. Battery life (hours)
25 & 125 15
1200 8
3600 mAh. Life at 125 ml/h (optional) up to 30 h

Pump alarms/alerts
Air in line / Up Occlusion, Down Occlusion, Near End Program, End Program / End of Infusion, Low Battery, End Battery, Pump Unattended, Door Open, System Error, Missing Set Key, with the use of flow sensor accessory

Ports
1 x RJ45 – for drop sensor
1 x RS232 (Male) – for PC communication

Additional specifications
Insulation class II, type CF
Ingress Protection: IP33
Linear Peristaltic mechanism